BIO 135 Virtual Lab
The Microscope
Return
to:
Unlabeled
Microscope Lab 2 Page
Microscope Usage
Notes
- Objective Lenses: A few of the microscopes may have 4 objective lenses,
with the largest labeled 100. Do not use this lens! It is an
oil-immersion lens and requires a drop of oil between the lens and the
slide.
- Ocular lens: Many
microscopes on the Leestown campus only have one ocular lens
- Fine and course focus controls: Some scopes have two separate controls
for this adjustment.
- Light Dimmer Switch: In addition to the main on/off switch to turn on the
light, you can also adjust the intensity of the light using the light dimmer
dial on the left side of the base (if the stage is facing you). If you
turn on the scope and you don't see any light (and it is plugged in!), try
turning this dial to increase the light intensity. It is usually best to
turn down the brightness somewhat while using low power, but you may need
to increase light intensity under higher magnification.
- Condenser: The condenser is situated directly beneath the stage,
and concentrates the light onto the specimen. It can be adjusted (raised
and lowered) with the control knob on the left side of the stage.
Adjusting the condenser allows you to improve the sharpness of the image.
If you have a bright spot (part of the field of view is brighter than the
rest), try lowering the condenser a bit. The condenser has some additional
controls:
- The condenser has an
internal iris that adjusts the size of the opening through which light
passes; it is controlled by a small black lever that slides back and
forth on the front of the condenser. Although it can be used to adjust
the amount of light reaching the specimen, its primary function is to
adjust the contrast of the image: opened up (more light) will produce
less contrast which may make fine details difficult to see; closed down
(less light) increases contrast, but can produce ghosting. (To just
control the brightness, use the light dimmer dial instead).
- There are also a pair of condenser alignment screws projecting from
the front of the condenser that align the condenser directly beneath the
objective lenses. You should not adjust these. If you notice that not all
of your field of view is lit (there may be a dark crescent toward one
edge), ask your instructor for assistance.
- You will be expected to demonstrate your ability to
correctly use the microscope with a Microscope Skills Quiz. You can
view the checklist
for this quiz so you will know what is expected of you.
This
page maintained by Bill Snyder . Last updated July 21, 2011.